Product

RMZN600-A&RMZN507-A Portable Dissolved Oxygen Meter

Model RMZN600-A&RMZN507-A
Output
Power supply
Application It can easily meet the monitoring demands of various water environment conditions such as sewage treatment, surface water, ocean and groundwater
Description RMZN507-A portable dissolved oxygen sensor is independently developed, using the international leading fluorescence lifetime technology, based on the principle of the quenching of active fluorescence by specific substances in physics. The significant advantages of the fluorescence method for measuring dissolved oxygen are that there is no oxygen consumption,no flow rate limit, no preheating, no electrolyte, no maintenance and frequent calibration, and the response time of the sensor (T90) is as low as 20 seconds, far better than the national standard requirements. RMZN600-A&RMZN507-A portable dissolved oxygen detector, with automatic temperature, pressure compensation, salinity manual input compensation function; Support automatic dissolved oxygen detection valueLock, support setting filtering mode;The dissolved oxygen detection has high accuracy and stability, and can be used in unstable water conditions such as aeration tank and sewage plant with higher effect. RMZN600-A multi-functional handheld instrument independently developed. Instrument with operating keyboard and backlight display, simple interface, comprehensive functions, simple operation, support most sensors plug and play, automatic identification, using the instrument can easily perform data measurement, data storage, parameter setting, sensor calibration and other functions,and support data export; The pursuit of high cost performance is our consistent pursuit.
  • Introduce
  • Datasheet

image.png

Product introduction

      RMZN507-A portable dissolved oxygen sensor is independently developed, using the international leading fluorescence lifetime technology, based on the principle of the quenching of active fluorescence by specific substances in physics. The significant advantages of the fluorescence method for measuring dissolved oxygen are that there is no oxygen consumption,no flow rate limit, no preheating, no electrolyte, no maintenance and frequent calibration, and the response time of the sensor (T90) is as low as 20 seconds, far better than the national standard requirements.

     RMZN600-A&RMZN507-A portable dissolved oxygen detector, with automatic temperature, pressure compensation, salinity manual input compensation function; Support automatic dissolved oxygen detection valueLock, support setting filtering mode;The dissolved oxygen detection has high accuracy and stability, and can be used in unstable water conditions such as aeration tank and sewage plant with higher effect.

    RMZN600-A multi-functional handheld instrument independently developed. Instrument with operating keyboard and backlight display, simple interface, comprehensive functions, simple operation, support most sensors plug and play, automatic identification, using the instrument can easily perform data measurement, data storage, parameter setting, sensor calibration and other functions,and support data export; The pursuit of high cost performance is our consistent pursuit.


Product operation

    The fluorescence dissolved oxygen sensor is based on the principle of quenching active fluorescence by specific substances in physics. The blue light emitted by a light-emitting diode (LED) is irradiated on the fluorescent substance on the inner surface of the fluorescent cap. The fluorescent substance on the inner surface is excited and emits red light. By detecting the phase difference between red light and blue light, and comparing with the internal calibration value, the concentration of oxygen molecules is calculated. The final value is automatically compensated by temperature and air pressure.

 Product features

◇High efficiency fluorescent film;

◇No oxygen consumption, no flow rate limit;

◇Real-time temperature and pressure compensation (connected to Runming handheld instrument);

◇Very fast response time, far better than the national standard requirements;

◇Metal body, split watertight design, better stability;

◇Built-in calibration parameters, factory has been accurately calibrated, easy to use on site;

◇Built in a number of algorithms, more efficient to adapt to the aeration tank, sewage plant water instability conditions (with Runming smart handheld instrument).

image.png

Product specification

Pseudolaric acid

Portable dissolved oxygen sensor

model number

RMZN507-A

Measurement principle

fluorimetry

Range                     0-20mg/L or 0-200% saturation

Precision

±1% or ±0.3mg/L(maximum)

Temperature range                     0-50℃

Response time

20s

Maximum pressure                     6bar

Protection level

IP68

junction              Equipped with RMZN600-A multi-function handheld meter

Probe size

φ26mm*175mm

Cable length                5 meters (default), customizable

Probe weight

204g

Calibration                 1 point or 2 point calibration

Shell material

316L

Pseudolaric acid

Multi-function handheld instrument

model number                    RMZN600-A

Probe weight

460g

Power source                2 18650 rechargeable batteries

Operating temperature

0-50℃

Storage temperature

-30~70℃

display

54.38x54.38mm LCD with backlight

Data storage

support

Air pressure compensation

Instrument built-in, automatic compensation 50~115kPa

Protection level                   IP67

Timed shutdown

support

Note: The above technical parameters are all data in the laboratory standard liquid environment.


Copyright © 2022-2026 Changsha Runming intelligent Technology Co., LTD

WhatsApp

+8615074952148